Validation of a chromatographic method for the assay of sulphamethoxazol and trimethoprinm in tablets
Type
Artículo de revista
Document language
EspañolPublication Date
1999Metadata
Show full item recordSummary
The results obtained during the standarization of a methodology by reverse high performance liquid chromatography for the assay of sulfamethoxazol and trimethoprim in tablets are present. For the assay we used a C18 reversed phase column and a mobile phase composed by methanol: phosphoric acid 0.01M (23:73). The wavelength for UV detection was fixed at 227 nm. Linearity was achieved between 20 and 140 mcg/mL and between 3 and 28 mcg/mL for sulfamethoxazol and trimethoprim repectively. The accuracy, precision and robutness were demostratedSummary
En el presente artículo se publican los resultados obtenidos durante el proceso de desarrollo y optimización de una metodología analítica por cromatografía líquida para la determinación cuantitativa de sulfametoxazol y trimetoprim en tabletas. El método utiliza una columna con octadecilsilano como fase estacionaria y una fase móvil compuesta por metanol: ácido fosfórico 0.01M ( 23:73 ) como fase móvil y detección espectrofotométrica a 227nm. El método luego de su validación demostró ser exacto, lineal y preciso dentro de un intervalo de concentraciones entre 20 y 140 mcg/mL para sulfametoxazol y entre 4 y 28 mcg/mL para trimetoprim.Collections