dc.description.abstractEn el presente trabajo se estudio y comparo dos sistemas dosimétricos utilizados en controles de calidad específicos para pacientes de Tratamiento de Radioterapia con Intensidad Modulada (Cabeza- Cuello y próstata) en la modalidad dinámica (slinding window): Películas EDR2 y matriz de diodos MapCHECK TM Model 1175. Las mediciones fueron realizadas en las instalaciones del área de radioterapia del Instituto Nacional de Cancerología-INC. Se trabajo con un grupo de 30 planeaciones correspondientes a pacientes con cáncer de cabeza-cuello y próstata. Se realizo la dosimetría absoluta al equipo Clinac IX, mostrando una variación en el rendimiento que no supera el 2% del valor nominal (1cGy/MU) para un haz de radiación de 6MV, con índice de calidad haz Q (20,10) de 0.664. Se examinaron los cálculos arrojados por los sistemas de planeación de tratamiento, mediante chequeos específicamente para haces de radiación fijados por el TECDOC-1583, los cuales mostraron una pequeña variación con respecto a la dosis entregada por el equipo de tratamiento dentro de los valores de tolerancia. Se evaluó la comparación entre la distribución de dosis planar entregada por los sistemas de planeación y los obtenidos tanto con los sistemas de detección EDR-2 y arreglo de diodos 2D (Map Check), para los métodos campo por campo y campos compuestos mediante los parámetros de índice gamma, diferencia de dosis porcentual (D%) y Distance to agreement (DTA), estos dos últimos con valores de tolerancia de 3% y 3mm respectivamente. Se obtuvo con el Map Check un valor de (%D) e índice gamma de 3,0±1,5 % y 96,41 ± 3.68% respectivamente. Para el caso de la película radiográfica EDR-2 se reportaron los valores [7]: (%D) inferior al 1,7% y un valor de índice gamma de 0,46 ± 0,14. Los resultados obtenidos tanto con el Map Check como con la película EDR- 2, están dentro de los valores de aceptación para control de calidad en IMRT, lo que los hace aptos a los dos sistemas de verificación para control de calidad en dicha técnica. Sin embargo en el caso del arreglo de diodos 2D Map Check se sugiere realizar al sistema (acelerador lineal, TPS) un completo comisionamiento previo a la realización de controles de calidad en IMRT. Por ultimo, se concluye de la comparación de los dos sistemas de verificación con los que se trabajo en este proyecto que la calidad del proceso de control de calidad en IMRT (especifico para paciente) realizado con película EDR-2 frente al arreglo de diodos 2D- Map Check, se mide dependiendo del proceso de revelado de la película, del scanner y del software de verificación utilizados para el análisis de datos.
dc.description.abstractAbstarct. In this paper we study and compare two dosimetric systems used in quality controls for patients with radiotherapy Intensity Modulated (Head-neck and prostate) in dynamic mode (slinding window): Movies EDR2 and diode array 1175. The Model TM MapCHECK Measurements were made in the area of ​​radiotherapy facilities Institute National Cancer-INC. We worked with a group of 30 plannings corresponding to patients with head and neck and prostate cancer. He performed the absolute dosimetry equipment Clinac IX, showing a variation in performance does not exceed 2% of the nominal value (1cGy / MU) for a radiation beam 6MV with beam quality index Q (20,10) of 0.664. They were examined by thrown stones treatment planning systems through checks specifically to do radiation set by the TECDOC-1583, which showed a small variation relative to the equipment delivered by the dose treatment within values tolerance. Comparison of the dose distribution was assessed by planar delivered planning systems and obtained both detection systems and EDR-2 2D arrangement of diodes (Map Check), for methods and fields field by field compounds by gamma parameters index, percentage difference dose (D%) and Distance to agreement (DTA), the latter two with values ​​of tolerance of 3% and 3mm respectively. It was obtained in the Map Check worth (% D) and gamma index 3.0 ± 1.5% and 96.41 ± 3.68% respectively. In the case of the radiographic film EDR-2 values ​​[7] reported: (% D) of less than 1.7% and a value of gamma index 0.46 ± 0.14. The results obtained with both the Map Check with the film as EDR- 2, are within the acceptance values ​​for IMRT QA, so the It makes them suitable both verification systems for quality control in this technique. However in the case of diode array 2D Map Check it is suggested to system (linac, TPS) complete prior to the completion of commissioning.
dc.format.mimetypeapplication/pdf
dc.language.isospa
dc.relation.ispartofUniversidad Nacional de Colombia Sede Bogotá Facultad de Ciencias
dc.relation.ispartofFacultad de Ciencias
dc.rightsDerechos reservados - Universidad Nacional de Colombia
dc.rights.urihttp://creativecommons.org/licenses/by-nc/4.0/
dc.subject.ddc51 Matemáticas / Mathematics
dc.subject.ddc53 Física / Physics
dc.subject.ddc61 Ciencias médicas; Medicina / Medicine and health
dc.subject.ddc62 Ingeniería y operaciones afines / Engineering
dc.titleAnálisis dosimétrico comparativo entre película radiográfica edr2 y arreglo de matriz de diodos en 2d para verificación de tratamientos de radioterapia de intensidad modulada (IMRT)
dc.typeTrabajo de grado - Maestría
dc.type.driverinfo:eu-repo/semantics/masterThesis
dc.type.versioninfo:eu-repo/semantics/acceptedVersion
dc.identifier.eprintshttp://bdigital.unal.edu.co/47070/
dc.description.degreelevelMaestría
dc.relation.referencesGutiérrez Díaz, Leina Patricia (2012) Análisis dosimétrico comparativo entre película radiográfica edr2 y arreglo de matriz de diodos en 2d para verificación de tratamientos de radioterapia de intensidad modulada (IMRT). Maestría thesis, Universidad Nacional de Colombia.
dc.rights.accessrightsinfo:eu-repo/semantics/openAccess
dc.subject.proposal2D arreglo de diodos
dc.subject.proposalPelicula EDR-2
dc.subject.proposalControl de calidad IMRT
dc.subject.proposal2D diode array
dc.subject.proposalEDR-2 film
dc.subject.proposalIMRT QA
dc.type.coarhttp://purl.org/coar/resource_type/c_bdcc
dc.type.coarversionhttp://purl.org/coar/version/c_ab4af688f83e57aa
dc.type.contentText
dc.type.redcolhttp://purl.org/redcol/resource_type/TM
oaire.accessrightshttp://purl.org/coar/access_right/c_abf2


Archivos en el documento

Thumbnail

Este documento aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del documento

Atribución-NoComercial 4.0 InternacionalEsta obra está bajo licencia internacional Creative Commons Reconocimiento-NoComercial 4.0.Este documento ha sido depositado por parte de el(los) autor(es) bajo la siguiente constancia de depósito