Analytical method validation for determination of total mercury in human blood by zeeman mercury spectrometer ra-915+ with the pyrolysis module pyro-915+
dc.contributor.author | Ramírez, Carolina | spa |
dc.contributor.author | Olarte, Ermel | spa |
dc.contributor.author | Tellez, Eliana | spa |
dc.contributor.author | Palma, Marien | spa |
dc.date.accessioned | 2019-06-29T09:20:33Z | spa |
dc.date.available | 2019-06-29T09:20:33Z | spa |
dc.date.issued | 2013 | spa |
dc.description.abstract | This article describes the methodology of the validation about the identification of total mercury in blood samples by Zeeman atomic absorption spectrophotometry RA-915+ with the module of Pyrolysis PYRO-915+, suitable for the direct determination of total mercury concentrations in various matrices, that unlike other atomization techniques (graphite furnace, cold vapor or hydride generator) does not require a sample pre-treatment. With the application of this validation, is provided the reliability in the results. Most of the validation parameters were subjected to analysis of variance ANOVA with a confidence level of 95%. It was obtained a detection limit of 2.16 µg/L, a quantitation limit of 5.54µg/L and linearity in the range of 5.54 µg/L to , in addition, the precision and accuracy was evaluated, obtaining favorable results in the analysis. The calculation of the expanded uncertainty for this method was 0.610 µg/L. | spa |
dc.description.abstract | Este artículo describe la metodología de validación para la técnica de identificación de mercurio total en muestras de sangre por espectrofotometría de absorción atómica Zeeman RA-915+ con el módulo de Pirólisis PYRO-915+, adecuada para la determinación directa de las concentraciones de mercurio total en varias matrices, que a diferencia de otras técnicas de atomización (horno de grafito, vapor frio, o generador de hidruros) no necesita un pre-tratamiento de la muestra. Con la aplicación de ésta validación en la realización de los análisis se brinda la confiabilidad necesaria en la obtención y emisión de los resultados. La mayoría de parámetros de la validación, se sometieron a análisis de varianza tipo ANOVA con un nivel de confianza del 95 %. Se obtuvo un límite de detección de 2.16 µg/L, un límite de cuantificación de 5.54 µg/L y una linealidad en el rango de 5.54 µg/L a 100 µg/L, adicionalmente, se evaluó exactitud y precisión, obteniendo resultados favorables para el análisis. El cálculo de la incertidumbre expandida para este método fue 0.610 µg/L. | spa |
dc.format.mimetype | application/pdf | spa |
dc.identifier.eprints | http://bdigital.unal.edu.co/43301/ | spa |
dc.identifier.uri | https://repositorio.unal.edu.co/handle/unal/49811 | |
dc.language.iso | spa | spa |
dc.publisher | Universidad Nacional de Colombia | spa |
dc.relation | http://revistas.unal.edu.co/index.php/rcolquim/article/view/45909 | spa |
dc.relation.ispartof | Universidad Nacional de Colombia Revistas electrónicas UN Revista Colombiana de Química | spa |
dc.relation.ispartof | Revista Colombiana de Química | spa |
dc.relation.ispartofseries | Revista Colombiana de Química; Vol. 42, núm. 3 (2013): Versión preliminar; 326-343 2357-3791 0120-2804 | |
dc.relation.references | Ramírez, Carolina and Olarte, Ermel and Tellez, Eliana and Palma, Marien (2013) Analytical method validation for determination of total mercury in human blood by zeeman mercury spectrometer ra-915+ with the pyrolysis module pyro-915+. Revista Colombiana de Química; Vol. 42, núm. 3 (2013): Versión preliminar; 326-343 2357-3791 0120-2804 . | spa |
dc.rights | Derechos reservados - Universidad Nacional de Colombia | spa |
dc.rights.accessrights | info:eu-repo/semantics/openAccess | spa |
dc.rights.license | Atribución-NoComercial 4.0 Internacional | spa |
dc.rights.uri | http://creativecommons.org/licenses/by-nc/4.0/ | spa |
dc.subject.proposal | mercurio | spa |
dc.subject.proposal | validación de métodos | spa |
dc.subject.proposal | espectrofotometría de absorción atómica | spa |
dc.subject.proposal | mercury | spa |
dc.subject.proposal | analytical methods validation | spa |
dc.subject.proposal | atomic absorption spectrophotometry | spa |
dc.title | Analytical method validation for determination of total mercury in human blood by zeeman mercury spectrometer ra-915+ with the pyrolysis module pyro-915+ | spa |
dc.type | Artículo de revista | spa |
dc.type.coar | http://purl.org/coar/resource_type/c_6501 | spa |
dc.type.coarversion | http://purl.org/coar/version/c_970fb48d4fbd8a85 | spa |
dc.type.content | Text | spa |
dc.type.driver | info:eu-repo/semantics/article | spa |
dc.type.redcol | http://purl.org/redcol/resource_type/ART | spa |
dc.type.version | info:eu-repo/semantics/publishedVersion | spa |
oaire.accessrights | http://purl.org/coar/access_right/c_abf2 | spa |
Archivos
Bloque original
1 - 1 de 1
Cargando...
- Nombre:
- 45909-222616-1-SM.pdf
- Tamaño:
- 998.34 KB
- Formato:
- Adobe Portable Document Format