Diseño e Implementación de una red de comunicación que permita recrear múltiples escenarios clínicos por medio de un módulo de control central hacia varios emuladores de entrenamiento neonatal

dc.contributorVarón Durán, Gloria Margaritaspa
dc.contributorMéndez Córdoba, Luis Carlosspa
dc.contributor.authorRamos Ballesteros, Edward Juliánspa
dc.date.accessioned2020-03-30T06:46:04Zspa
dc.date.available2020-03-30T06:46:04Zspa
dc.date.issued2019spa
dc.description.abstractEste trabajo presenta el diseño e implementación de una red de emuladores neonatales, compuesta por un Módulo de Control Central y tres Emuladores. La red le permite a los docentes del área de la salud crear escenarios clínicos por medio del Módulo de Control Central. Los escenarios son en general, patologías representadas por algunos valores de ciertos signos vitales, en este caso, se involucra lo concerniente al sistema respiratorio de pacientes neonatales. Luego de crear los escenarios, estos son transmitidos hasta el Emulador escogido por el docente mediante una interfaz gráfica. Los estudiantes en práctica observan los valores y señales de los signos vitales tanto en la parte física (Maniquí del Emulador) como en la parte gráfica (Interfaz del Emulador), y pueden realizar algún tipo de acción sobre el Emulador dependiendo de los conocimientos previamente adquiridos en sus estudios universitarios. Con base en las acciones llevadas a cabo por el estudiante en práctica, se genera un proceso de respuesta con el objetivo de indicarle al docente los resultados obtenidos. Después, el docente compara el escenario clínico previamente programado con su respectiva respuesta. De esta forma, se evalúa a los estudiantes en práctica y se estimula el aprendizaje de los mismos. Después de ejecutar todos los respectivos procesos de diseño, implementación y optimización, la red cuenta con un rango de 9 metros de cobertura, con la posibilidad de aumentarlo en el caso donde solo uno de los componentes del Emulador se aleje, además una de esas partes; Maniquí del Emulador, cuenta con una autonomía de 44.2 horas aproximadamente gracias a la capacidad de la batería y a los componentes de bajo consumo. Con respecto a los procesos de comunicación principales, en promedio la programación del Escenario clínico dura 2.2 segundos y la Respuesta dura 2.3 segundos, estos tiempos son adecuados para la aplicación. En cuanto a la programación, los algoritmos están debidamente parametrizados y también se cuenta con 27 bits libres dentro de la trama total. Los dos aspectos anteriores, hacen más fácil futuras transiciones en el momento de adicionar herramientas a la red o agregar características a los escenarios clínicos, ya sea del sistema respiratorio u otros sistemas, sin la necesidad de modificar el mecanismo de transmisión o recepción. Del mismo modo, se crearon, aplicaron y programaron varios procesos particulares de un sistema de comunicación en red, los cuales en conjunto convierten a la red de emuladores neonatales en un sistema confiable. Todos los aspectos mencionados anteriormente, generan una red robusta con herramientas apropiadas para cumplir los objetivos establecidos y brindar un instrumento tecnológico de aprendizaje, por medio del cual, el docente comparte simultáneamente sus conocimientos a varios estudiantes en práctica. Así mismo, el sistema es adecuado considerando las características físicas del espacio donde va estar ubicado y las especificaciones de desempeño enmarcadas en la idea de implementar un taller práctico.spa
dc.description.abstractAbstract: The following project presents the design and implementation of a network of neonatal emulators, composed by a Central Control Module and three emulators. The network allows health areas professors to create clinical scenarios using the Central Control Module tool. These scenarios commonly are, pathologies represented by the values of certain vital signs, in this case, the respiratory system of neonatal patients. After creating the scenarios, these are transmitted to the Emulator chosen by the professor through a graphic interface. The students in practice observe the values and signals of the vital signs both in the physical part (emulator mannequin) and the graphical part (emulator interface), and can make some type of action on the emulator based on their previous attained knowledge from their university studies. The actions performed by the student in practice, generates a response process with the purpose of indicating the professor the results obtained. Afterwards, the teacher compares the clinical scenario previously programmed with the corresponding result. In this way, the students in practice are tested and their learning is stimulated. After executing all the respective procedures of design, implementing and optimizing, the network has a coverage of 9 meters, with the possibility of increasing it in case that only one component of the Emulator is moved away. The Emulator Mannequin, counts with an autonomy of approximately 44.2 hours due to the capacity of the battery and to the components of low consumption. Regarding the communication processes main, in average the programming of the clinical scenarios lasts 2.2 seconds and the answer takes 2.3 seconds, these timings are suitable for the application. As for the programming, the algorithms are duly parameterized and they also count with 27 bits within the total weft. The two previous aspects make future transitions easier in the moment of adding tools to the network or to adding characteristics to the clinical scenarios, either the respiratory system or other systems, without having to modify the mechanism of transmission or reception. In the same way, various particular processes of a network communication system were created, applied, and programmed, which altogether convert the network of neonatal emulators in a trusting system. All aspects previously mentioned, generate a robust network with tools appropriate to accomplish the objectives stablished and provide a technological instrument for learning, whereby, the professors can share their knowledge simultaneously to different students in practice. Likewise, the system is adequate considering the physical characteristics of the room where it will be located and the specifications of performance framed in the idea of implementing a practical workshop.spa
dc.description.degreelevelMaestríaspa
dc.format.mimetypeapplication/pdfspa
dc.identifier.eprintshttp://bdigital.unal.edu.co/74958/spa
dc.identifier.urihttps://repositorio.unal.edu.co/handle/unal/77308
dc.language.isospaspa
dc.relation.haspart02 Bibliotecología y ciencias de la información / Library and information sciencesspa
dc.relation.haspart61 Ciencias médicas; Medicina / Medicine and healthspa
dc.relation.ispartofUniversidad Nacional de Colombia Sede Bogotá Facultad de Ingeniería Departamento de Ingeniería Eléctrica y Electrónica Ingeniería Electrónicaspa
dc.relation.ispartofIngeniería Electrónicaspa
dc.relation.referencesRamos Ballesteros, Edward Julián (2019) Diseño e Implementación de una red de comunicación que permita recrear múltiples escenarios clínicos por medio de un módulo de control central hacia varios emuladores de entrenamiento neonatal. Maestría thesis, Universidad Nacional de Colombia - Sede Bogotá.spa
dc.rightsDerechos reservados - Universidad Nacional de Colombiaspa
dc.rights.accessrightsinfo:eu-repo/semantics/openAccessspa
dc.rights.licenseAtribución-NoComercial 4.0 Internacionalspa
dc.rights.urihttp://creativecommons.org/licenses/by-nc/4.0/spa
dc.subject.proposalEmulador Neonatalspa
dc.subject.proposalEscenarios Clínicosspa
dc.subject.proposalRadio Frecuencia (RF)spa
dc.subject.proposalIdentificación por Radio Frecuencia (RFID)spa
dc.subject.proposalRed de Comunicaciónspa
dc.subject.proposalSistema Respiratoriospa
dc.subject.proposalNeonatal emulatorspa
dc.subject.proposalClinical scenariosspa
dc.subject.proposalRadio Frequency (RF)spa
dc.subject.proposalRadio Frequency Identification (RFID)spa
dc.subject.proposalCommunication Networkspa
dc.subject.proposalRespiratory systemspa
dc.titleDiseño e Implementación de una red de comunicación que permita recrear múltiples escenarios clínicos por medio de un módulo de control central hacia varios emuladores de entrenamiento neonatalspa
dc.typeTrabajo de grado - Maestríaspa
dc.type.coarhttp://purl.org/coar/resource_type/c_bdccspa
dc.type.coarversionhttp://purl.org/coar/version/c_ab4af688f83e57aaspa
dc.type.contentTextspa
dc.type.driverinfo:eu-repo/semantics/masterThesisspa
dc.type.redcolhttp://purl.org/redcol/resource_type/TMspa
dc.type.versioninfo:eu-repo/semantics/acceptedVersionspa
oaire.accessrightshttp://purl.org/coar/access_right/c_abf2spa

Archivos

Bloque original

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
Red de Emuladores Neonatales - Julián Ramos.pdf
Tamaño:
1.46 MB
Formato:
Adobe Portable Document Format