Evaluation of materials and micro shot peening to increase resistance to failure due to cyclic fatigue in rotary endodontic files
Advisor
Type
Trabajo de grado - Doctorado
Document language
InglésPublication Date
2022-06-03Metadata
Show full item recordAbstract
This thesis tackles the problem of endodontic fi les fractures from two views; In the fi rst instance, alternative materials to the classic Nickel-Titanium alloy such as Titanium-Niobium and Titanium-Molybdenum were analyzed. This analysis found that Titanium-Niobium could be proposed as an alternative for manufacturing endodontic files. Secondly, the shot peening procedure was proposed and analyzed to increase the resistance to fatigue fracture both in the proposed alternative alloys and in Nickel-Titanium fi les, fi nding a signi cant increase in the resistance to a fatigue fracture in the fi les and wires subjected to shot peening.Summary
Esta tesis aborda el problema de las fracturas de las limas endodonticas desde dos puntos de vista; En primera instancia se analizaron materiales alternativos a la clasica aleacion de Niquel-Titanio como son el Titanio-Niobio y el Titanio-Molibdeno. Este analisis encontro que el Titanio-Niobio se puede proponer como una alternativa para la fabricacion de limas endodonticas. En segundo lugar, se propuso y analizado el procedimiento de shot peening para aumentar la resistencia a la fractura por fatiga tanto en las aleaciones alternativas propuestas como en las limas de Niquel-Titanio, encontrando un aumento signi cativo de la resistencia a la fractura por fatiga en las limas y alambres sometidos al shot peening. (Texto tomado de la fuente)Keywords
Physical description
ilustraciones, fotografías, graficas